Kirby, Alanah and Rowe, P J (2002) Ultrasound Measurement of Femoral Anteversion and Tibial Torsion. Project Report. Scottish Executive, Final Report submitted to the Chief Scientist Office. Edinburgh: Scottish Executive..
| PDF - Requires a PDF viewer such as GSview, Xpdf or Adobe Acrobat Reader 75Kb |
Official URL: http://www.cso.scot.nhs.uk/Publications/ExecSumms/OctNov02/czg.4.1.6.doc
Abstract
This project aimed to establish a method of measuring femoral anteversion and tibial torsion using real time ultrasound images and a new system of electronic inclinometry. The inclinometry system was carefully calibrated and proved accurate to within 1 degree and precise to with 0.2 of a degree. Individual inclinometers were attached firmly to both a 5MHz and a 7.5MHz linear array ultrasonic transducer using a specially produced clamp which gripped the front and rear of the probe head. Two low profile flexible switches linked by wire to a data collection computer were used to start and stop recording of the output of the inclinometers at 50Hz
| Item Type: | Monograph (Project Report) |
|---|---|
| ID Code: | 2856 |
| Deposited On: | 03 Jul 2012 16:16 |
| Last Modified: | 03 Jul 2012 16:25 |
Repository Staff Only: item control page